The Rig Lizard Series
Specifically designed to provide maximum grip in difficult situations, with several options for dry, oil or mud work.
Replacement sensor for gas detectors of the series: MicroClip XL for measurement of oxygen concentration in a range of 0-30 Vol Oxygen within a normal temperature range and conditions. Manufacturer CitiCell, Brand BW
They are designed to operate in a wide range of harsh environments and conditions. However, it is important that exposure to high concentrations of solvent vapors be avoided, both during storage and during instrument installation and operation.
When using sensors with printed circuit boards (PCBs), degreasing agents must be used before the sensor is fitted. Do not glue directly on or near the CiTiceL as the solvent may cause cracks in the plastic.
Toxicity
Toxic gases at TLV levels will not have any cross-sensitivity effects on oxygen CiTiceLs. At very high levels (i.e. percentage
levels), highly oxidizing gases (e.g., ozone and chlorine) will interfere with the measurement of their oxygen equivalent, but
most other commonly occurring gases will have no effect.
Compatibility
Gas Alert MicroClip | After Software Update it is Compatible |
Gas Alert Microclip XL | Compatible |
Gas Alert Microclip XT | Compatible |
Gas Alert Microclip X3 | Not Compatible See SR-X3V |
Environment or surroundings
Operating Temperature Range | -20°C to +50°C |
Storage temperature | 0°C to 20°C |
compensation temperature | Refer to OP-5 |
Atmospheric Pressure Range |
Atmosphere+10% |
Atmospheric Pressure Coefficient | 0.02% Signal/mbar |
Humidity Range | 0-99% RH uncondensed |
Time of life | |
Loss of Sensitivity | <5% loss of precision per year |
Expected Life Time in Operation | Two years in the air |
Storage Life | 6 Months in its Container |